IM 11G06A01-02EN
5-54
5. SCREENS
1st Edition : Oct. 27, 2023-00
5.6.2
Loop check
Allows output of the value set at the current output terminal (AO).
1 Tap the icon
on the measurement screen.
The menu screen appears.
2 Touch "Maintenance".
The maintenance menu will appear.
3 Select "Loop check".
The Loop check screen will appear.
4 Tap "Test validity AO".
A dialog box for multiple selections will appear.
Select the AOs you wish to test (multiple selections allowed).
5 Tap "AOx test output".
A keyboard dialog appears.
Enter the current value you wish to output.
(x: 1 to 4)
6 If you change "Test validity AO" or "AO○ Test Output", the background of
the changed item will turn blue and the
in the upper right corner of
the screen will change to a
.
7 Tap the
to confirm the change and the set current value will be output
from the enabled AO
l
Description of valve selection
Test validity AO.
Setting the current output terminal to be output
AOx test output
Setting the value to be output to the current output terminal
(gas name) Span: Valve status during span calibration of each component
(x: 1 to 4)
l
Setting range
AOx test output
2.4 to 21.6 mA (default 4.0 mA)
Summary of Contents for IR800G
Page 17: ...Blank Page...
Page 47: ...Blank Page...
Page 121: ...Blank Page...
Page 123: ......
Page 125: ......
Page 127: ...Blank Page...