IM 11G06A01-02EN
5-47
5. SCREENS
1st Edition : Oct. 27, 2023-00
5.4.5
Automatic Validation
Use this option when you want to perform automatic validation immediately, regardless of the
initial start or cycle of the automatic validation setting. Even if the "On/Off" setting of the automatic
validation setting is "Off", the automatic validation will be executed according to the contents of
the automatic validation setting. (See 5.1.5 Setting Automatic Validation for details on automatic
validation settings.)
l
Starting on screen
(1) Tap
on the Home/Main screen > Calibration
(2) Tap
to select Automatic validation.
(3)
Confirm the validation contents (channels to be automatically validated
and calibration range) on the screen that appears. If there is no problem,
tap the [Start validation]. The automatic validation starts and the
measurement screen is displayed.
(4)
indicating that validation is in progress is displayed at the top of the
screen in the measurement screen, and the Ch (component) subject to
automatic validation is switched to the calibration range.
During zero validation, the
is displayed on the Ch (component) to
be automatically validated, and zero validation is executed at the same
time. During span validation, span validation is executed one by one in
the order of Ch (components)
is displayed on the Ch (component)
under span validation.
(5) When all validations are completed, the
blinks during the
substitution time. The range returns to the range before validation. When
the displacement time has elapsed, the
goes off and the instrument
returns to the measurement state.