IM 11G06A01-02EN
5-44
5. SCREENS
1st Edition : Oct. 27, 2023-00
l
Contact output during automatic calibration
The contact output during automatic calibration varies as shown in Figure 5.14 and Figure 5.15
according to the gas flow channel setting in the automatic calibration setup. (See 5.2.3 Contact
output setting for the functions assigned to the contact outputs and 5.1.2 Auto Calibration Setting
for the gas flow channel.)
for sample switching
for zero calibration
for CH1 span calibration
for CHx span calibration
Calibration start
End of calibration
Figure 5.14
Gas flow channel setting: Analyzer
for system calibration
for zero calibration
for CH1 span calibration
for Chx span calibration
Calibration start
End of calibration
The contact output for "sample switching" remains ON.
Figure 5.15
Gas flow channel setting: Sampling system
5.4.4
Auto Zero Calibration
This is used when you want to perform a auto zero calibration immediately, regardless of the
initial start or cycle of the zero calibration setting. Even if the "On/Off" setting of the auto zero
calibration setting is "Off", it will be executed according to the contents of the auto zero calibration
setting. (See 5.1.3 Auto zero calibration setting for details on the Simple Zero Calibration
setting.)