IM 11G06A01-02EN
2-3
2. SPECIFICATIONS
1st Edition : Oct. 27, 2023-00
Safety, EMC and RoHS conformity standards
Safety conformity standards:
CE, UKCA EN 61010-1,
EN IEC 61010-2-030
UL
UL 61010-1,
UL 61010-2-030
CSA CAN/CSA-C22.2
No.61010-1, 61010-2-030
GB GB30439
Installation altitude: 2000 m or less
Installation category;
(IEC 61010)II (Note 1)
Pollution degree;
(IEC 61010); 2 (Note 2)
Note1: Installation category, so called overvoltage category,
specifies impulse withstanding voltage. Category
II overvoltage applies to equipment intended to be
powered from the building wiring.
Note2: Pollution degree indicates the degree of existence of
solid, liquid, gas or other inclusions which may reduce
dielectric strength. Degree 2 indicates the normal
indoor environment.
EMC:
CE, UKCA EN61326-1 Class A,
Table 2 (For use in
industrial locations)
EN61326-2-3,
EN61000-3-2,
EN IEC 61000-3-2,
EN61000-3-3
RCM
EN61326-1 CLASS
A, Table2
Note: · This instrument is a Class A product, and it is designed
for use in the industrial environment. Please use this
instrument in the industrial environment only.
· Influence of immunity environment (Criteria A): Output
shift is specified within ±15% of F.S.
Environmental regulation
RoHS; EN IEC 63000
Information of the WEEE Directive
This product is purposely designed to be used in a
large scale fixed installations only and, therefore,
is out of scope of the WEEE Directive. The WEEE
Directive does not apply. The WEEE Directive is
only valid in the EU/UK.
REACH; Regulation EC 1907/2006
Standard Functions
Output signal hold:
Measured value output can be
held during calibration or operation
by automatic or on-screen
operation, or by remote execution
instructions. The output at hold
can be selected from the previous
value and the set value.
Range changeover:
The output range of measured
values can be changed
automatically or manually. Range
changeover can also be turned off.
Range identification signal:
When using the range changeover
function, it is possible to output
whether the low or high range is
being used.
Blowback: Can open/close the contact output
for blowback by a scheduled cycle
or by an execution command.
Blowback can set the blowback
time and the gas displacement time
after blowback (output hold time).
Auto calibration:
Can be calibrated automatically
by scheduled cycles or execution
commands. If the calibration
coefficient is outside the normal
range, an alarm is issued. The
alarm can be assigned to a contact
output.
Auto zero calibration:
Separate from the schedule set for
Auto calibration, Auto calibration
for Zero Point only can be set. If the
timing of Auto calibration and the
Auto zero calibration are to occur at
the same time, the Auto calibration
takes precedence.
Auto validation:
-V is selected, the automatic
validation function is available.
The automatic validation function
flows the calibration gas through
the measurement line and records
the measured value according to
a scheduled cycle or execution
command. If the measured value
deviates from the set threshold
value, an alarm can be issued and
the alarm can be assigned to a
contact output. This function allows
the user to verify the normality of
the instrument.
Contact output during auto-calibration/validation:
Displays the status of automatic
calibration and validation. Status
can be assigned to a contact output.
High/low limit alarm:
Upper or lower thresholds can
be set and alarmed if exceeded.
Alarms can be assigned to contact
outputs.