IM 11G06A01-02EN
5-22
5. SCREENS
1st Edition : Oct. 27, 2023-00
Example of average behavior
When the average time is set to 1 hour
Reset
Setting time
Time
Average
value
Figure 5.8
● Calculated by averaging the data acquired once every 30 seconds for 1 hour (set time).
(Data acquired once every 30 seconds is a 30-second average of data acquired at 500 ms.)
● The average value is output every 30 seconds from the current time up to 1 hour (set time)
ago.
● At the time of reset, all values prior to the current time are calculated as zero. Therefore,
accurate average values will not be available until one hour after the reset.
5.2.2
mA output settings
In the mA current output setting, set the current output pin assignment, hold, response time,
output limit, and other settings related to current output.
(1) Tap
on the Home/Main screen > Menu
(2) Tap [Setting] > Settings
(3) Tap [mA output settings]
(4) Tap each item to set.