IM 11G06A01-02EN
5-9
5. SCREENS
1st Edition : Oct. 27, 2023-00
5.1.4
Validation channel setting
Set the range and behavior of alarms during automatic validation. There are two items to set:
validation alarm range and automatic validation
This setting screen is displayed only when the code "-V" in the Automatic Validation is specified.
(1) Tap
on the Home/Main screen > Menu
(2) Tap [Cal/Alarm Setting] > Cal/Alarm settings
(3) Tap [Validation Ch.setting].
For each channel, the components (the text is displayed in gray and
cannot be changed), the validation alarm range, and the auto validation
items and their set values are displayed.
(4) Tap each item to set.
l
Setting Validation Alarm Ranges
During automatic validation, you can check the validity of the measurement and whether
calibration is necessary or unnecessary by confirming that the difference between the flowing
calibration gas (zero/span) and the actual measured value is within the validation alarm range.
Touch the Validation Alarm Range item and enter a value in the keyboard dialog that appears.
During the automatic validation, if the measured value exceeds this validation alarm range, a
validation alarm is triggered.
Alarm range (%FS)
Alarm range (%FS)
Reference value
(Zero/Span)
l
Setting Auto Validation
Ch (component) settings for automatic verification. First, touch the "Automatic Validation" item.
In the selection dialog that appears, select the automatic validation (enable/disable) for the target
Ch (component).