IM 11G06A01-02EN
3-12
3. INSTALLATION
1st Edition : Oct. 27, 2023-00
The converter receives a contact signal and performs the configured function.
Cable specifications:
Each CH requires two cores for this wiring.
Select the number of cores according to the number of contacts to be used.
Wiring instructions:
(1) The terminal screw of the converter is an M3 screw. The tightening torque is 0.6 N•m.
Use crimp terminals compatible with these screws to terminate the cable.
(2) Resistance or voltage values determine the “ON/OFF” of this contact input.
Switches must meet the conditions shown in the table below.
Table 3.1
ON/OFF” identification of contact input
CLOSE
OPEN
Resistance
200 Ω or less
100 kΩ or more
Voltage value
-1 to 1 VDC
4.5 to 25 VDC
l
Terminal block 2
Terminal block 2
Shield must be connected to FG terminal
analog input
(O
2
analyzer: available when -2
is selected)
1 AI+
2
FG
RS-485
(Digital communication:
Available when -R is selected)
3 AI-
4 GND
Unused pins (not connectable)
5 N.C.
6
B-
If termination is required,
connect A+ and TERM.
7 N.C.
8
A+
Analog output CH1
9 AO1+ 10 TERM
unused pin
(No connection)
11 AO1- 12 N.C.
Analog output CH2
13 AO2+ 14 N.C.
15 AO2- 16 N.C.
Analog output CH3
17 AO3+ 18 N.C.
19 AO3- 20 N.C.
Analog output CH4
21 AO4+ 22 N.C.
23 AO4-
M3 screw terminal
Figure 3.9
Terminal block 2: analog input/analog output/RS-485
This terminal block has the following three functions.
● Analog input ( O
2
analyzer: available when “-1” or “-2” is selected)
When using an external oxygen meter to input the O
2
detector, use this terminal for current input
(4-20 mA). Set the 20 mA oxygen concentration to 100% in the external oxygen meter setting.
Cable specifications:
Use two-core cables for this wiring.
Wiring instructions:
(1) Use an M3 screw for the terminal screw of the converter. The tightening torque is 0.6 N•m.
Use crimp terminals compatible with these screws to terminate the cable.
(2) Do not wire with reversed polarity. Wire carefully to avoid mistaking the “+” and “-” polarity.
Summary of Contents for IR800G
Page 17: ...Blank Page...
Page 47: ...Blank Page...
Page 121: ...Blank Page...
Page 123: ......
Page 125: ......
Page 127: ...Blank Page...