IM 11G06A01-02EN
2-4
2. SPECIFICATIONS
1st Edition : Oct. 27, 2023-00
Instrument error contact output:
If an instrument error occurs,
an alarm is issued. The alarm is
assigned to a contact output.
Calibration error contact output:
If calibration is not performed
properly, an alarm is issued. The
alarm is assigned to a contact output
Optional Functions
IR800G, IR810G common
/U: unit conversion (mg/m
3
, g/m
3
)
Changes the units for instantaneous values,
O
2
conversion, and average values of NO, SO
2
, and CO to mg/m
3
or g/m
3
. If the measurement
range is set to vol%, no conversion is made.
See Table 2.3 for the corresponding values.
Table 2.3
ppm-mg/m
3
range
ppm range
Corresponding range in mg/m
3
NO
SO
2
CO
0-50 ppm
0-65.0 mg/m
3
0-140 mg/m
3
0-60.0 mg/m
3
0-100 ppm
0-130 mg/m
3
0-280 mg/m
3
0-125 mg/m
3
0-200 ppm
0-260 mg/m
3
0-570 mg/m
3
0-250 mg/m
3
0-250 ppm
0-325 mg/m
3
0-700 mg/m
3
0-300 mg/m
3
0-300 ppm
0-400 mg/m
3
0-850 mg/m
3
0-375 mg/m
3
0-500 ppm
0-650 mg/m
3
0-1400 mg/m
3
0-600 mg/m
3
0-1000 ppm
0-1300 mg/m
3
0-2800 mg/m
3
0-1250 mg/m
3
0-2000 ppm
0-2600 mg/m
3
0-5600 mg/m
3
0-2500 mg/m
3
0-2500 ppm
0-3300 mg/m
3
0-7100 mg/m
3
0-3000 mg/m
3
0-3000 ppm
0-4000 mg/m
3
0-8500 mg/m
3
0-3750 mg/m
3
0-5000 ppm
0-6600 mg/m
3
0-14.00 g/m
3
0-6250 mg/m
3
/P: Analyzer internal purge port:
For the following cases, /P should be
selected.
• When the sample gas contains combustible gas
•
When corrosive gases are present in the
atmosphere at the installation site.
• When the atmosphere at the installation site
contains the same gases as the measured
component.
Use dry N
2
or instrument air free of dust or
mist as the purge gas. The required flow
rate is approximately 1 L/min.
/A: CO Peak Alarm:
An alarm is issued when the number of
times the measured concentration of CO
peaks above the upper limit exceeds
the set value. The alarm is assigned to a
contact output.
/K: O
2
correction; Instantaneous O
2
correction
values and O
2
correction average values
of NO, SO
2
, and CO can be calculated and
output.
/NX: NO on the measurement screen is displayed as
NOx:
NO is displayed as NOx. /U. When used in
conjunction with /U, the units are converted
as NO
/PR: Pressure regulator valve
(Set for sample gas/ reference gas):
Includes one pressure regulator valve
for reference gas and one for sample
gas.
Pressure regulator valve for sample gas
Part Number: K8019GA
Inlet pressure: 7 to 18 kPa
Outlet pressure: 5±0.05 kPa
Gas contact material: Stainless
steel
Pipe connection: Ø6 mm/Ø4 mm
PTFE tube
Pressure regulator valve specifications for
reference gas
Part Number: K8019GB
Inlet pressure: 7 to 18 kPa
Outlet pressure: 5±0.05 kPa
Gas contact material: aluminum
Pipe connection:
Ø
6 mm/
Ø
4 mm
PTFE tube
IR810G only
/CG1~/CG4: Cable gland for wiring:
Cable gland for wiring is
included. If you do not provide
your own, please select the
required number of cable
glands.
/RP IR202-B: Sheet metal for wall-mounted
replacement:
Select this option when installing
the same dimensions as
IR202-B.
Performance
NO/SO
2
/CO/CO
2
/CH
4
Repeatability: ± 0.5% F.S.
(±1% F.S. when the optional
range is included)
Linearity: ±1% F.S.
Zero drift: ±1.0% F.S./week
(±2% F.S./week when the
optional range is included)
Span drift: ± 2.0% F.S./week
Response time (90% F.S. response): 30 sec. or
less
Interference: ± 2.0% F.S. (for Sample gas