IM 11G06A01-02EN
4-8
4. OPERATION
1st Edition : Oct. 27, 2023-00
(5) Alarms
Problems with the equipment will trigger an alarm. There are two types of alarms, Fault
or
Warning
. When multiple alarms occur, the highest priority alarm is displayed at the top of the
screen (representative alarm).
Severity Priority >
For some alarms, a text icon indicating the contents of the alarm is displayed in a region for each
channel (individual alarms).
Individual alarm
Character Icons
Main alarm
Individual alarm
Figure 4.9
Zero calibration alarm
Span calibration alarm
Automatic calibration alarm
Calibration Stability Alarm
Zero Calibration Caution
Span Calibration Caution
Zero-validation alarm
Span Validation Alarm
High high limit alarm
High limit alarm
Low limit alarm
Low low limit alarm
See 7.1 Alarms for the types of alarms displayed and their contents.
l
Screen operation when an alarm occurs
The screen operation in the event of an alarm is shown below. See 7.1 Alarms.
Measurement screen
or
扌
or
Alarm list window
Tap the corresponding
error
扌
Alarm detail window
Figure 4.10