IM 11G06A01-02EN
5-12
5. SCREENS
1st Edition : Oct. 27, 2023-00
l
Example
On/Off
On
First start day
Sun.
First start time
12:00
Interval
1
Interval unit
day
Flow time
zero
350 sec.
Ch1 span
350 sec.
Ch2 span
350 sec.
Ch3 span
350 sec.
Ch4 span
300 sec.
Ch5 span
300 sec.
Replacement
300 sec.
When automatic validation is performed with the above settings:
Flow time
:Auto validation in progress
Sunday
12
:
00
Interval
Zero
Validation
350 sec.
Ch1
Span validation
Ch2
Span validation
Monday
12
:
00
Tuesday
12
:
00
350 sec.
350 sec.
350sec.
300sec.
300 sec.
Ch4
Span validation
Ch3
Span validation
Ch5
Span validation
300 sec.
Replacement
(This is an example when Ch1 to Ch5 are set to "Enable" in "5.1.5 Automatic Validation Setting" and the oxygen analyzer is a
built-in oxygen analyzer.)
Figure 5.4
l
Setting range
Interval
1 to 99 hours or 1 to 40 days (default 7 days)
Flow time (zero, replacement time)
60 to 900 seconds (default 180 seconds)
Flow time (each span gas)
5 to 900 seconds (default 180 seconds)
CAUTION
• When automatic validation starts, the measurement screen is automatically displayed.
• After the power is turned off (including power failure, etc.) and then turned on again, the next
automatic validation is executed at the time set for the start time, and then at intervals of
cycles.
• Regardless of whether automatic validation is On or Off, it can be performed from the
screen or by remote start.