5. SCREENS
1st Edition : Oct. 27, 2023-00
5. SCREENS
5.1
Calibration/Alarm setting
5.1.1
Calibration channel setting
The Calibration Channel setting is used to set the calibration range, span gas concentration, and
operation. There are four items to be set: Calibration Range, Span Gas Concentration, Manual
Zero Calibration, and Auto Calibration/Auto Zero Calibration.
(1) On Main screen, touch
to go to the Menu.
Menu is displayed.
(2) Touch [Cal/Alarm Setting]
Calibration/Alarm setting is displayed.
(3) Touch [Calibration Ch. setting]
For each channel, the component (text is displayed in gray and cannot
be changed), calibration range, span gas concentration, manual zero
calibration, and Auto Calibration/Auto Zero Calibration items and their set
values are displayed.
(4) Touch each item to set.
l
Items to set
Calibration Range
Setting the range to be calibrated during calibration
Span gas concentration
Set the span gas concentration to be used during calibration
Manual Zero Calibration
Setting whether to calibrate other measurement components at
the same time or individually during manual zero calibration
Auto Calibration/Auto Zero Calibration Ch (component) setting for automatic calibration/auto
zero calibration