IM 11G06A01-02EN
5-21
5. SCREENS
1st Edition : Oct. 27, 2023-00
l
Setting range
Average time:
1 to 4 h (when the unit is hour) or 1 to 59 min (when the unit is minute) (default 1 h)
Range setting
(1) Tap the
on the channel setting screen to display the range setting item for the channel
you wish to change, and then tap that item.
(2) The range setting screen appears.
Tap the
to display the item whose settings you wish to change, then touch the item.
Use the selection dialog that appears or the keyboard dialog to change the setting. (Lower
and upper ranges cannot be changed.)
CAUTION
• Ch x Range setting is displayed only when no verification option is specified for that Ch.
• If the range is changed, the upper and lower limits of the alarm may exceed the range, but
this does not result in a setting error. In such cases, check the upper and lower alarm limits.
Alarm setting for details on alarm settings.)
Range Setting Limits
(1) The relationship between Range 1 value, Range 2 value, lower limit value, and upper limit
value shall be set to satisfy the following equation.
Lower limit range ≤ Range 1< Range 2 ≤ Upper limit range
(2) The maximum number of digits that can be entered in the range is 4. Decide the number
of decimal places considering that the maximum value varies depending on the number of
decimal places.
Example Number of decimal places = 2: Maximum value = 99.99
(3) The number of ranges can be 1 or 2; if 1, the Range 2 setting is invalid.
CAUTION
• Lower and upper limits are specified by the model and suffix code at the time of purchase
and cannot be changed.
• Perform zero and span calibration for each range setting change.
Range Switchover
Manual Operate by switching on the measurement screen. (See 4.4 General operation of
Remote Assign "Ch x Remote Range" of the target Ch (component) to the contact input. (x is 1
to 5) Assign Range 1 when there is an input to that contact and Range 2 when there is
no input. (For contact input settings, refer to 5.2.4
Auto
Automatically switches from Range 1 to Range 2 when the measured concentration
value exceeds 90% of Range 1, and from Range 2 to Range 1 when the measured
value is less than 80% of Range 1.