IM 11G06A01-02EN
5-6
5. SCREENS
1st Edition : Oct. 27, 2023-00
CAUTION
• The flow time setting screen will only display the currently used Ch.
• The replacement time is the extended time to hold the output signal after the calibration is
completed. It is extended only when the hold setting is On. The replacement time set here is
also the hold extension time during manual calibration.
• Flow time setting should be long enough to allow the concentration value to stabilize.
• When a mixed gas containing multiple components is used as the calibration gas, the flow
time setting can be set to 5 seconds, the shortest possible time, because the continuous
calibration of those components eliminates the need for a second and subsequent flow time
setting. (The assignment of components to Ch is fixed and cannot be changed.)
l
Example of the setting
On/Off
On
First start day
Sunday
First start time
12:00
Interval
1
Interval unit
day
Flow time
Zero
350 sec.
Ch1 span
350 sec.
Ch2 span
350 sec.
Ch3 span
350 sec.
Ch4 span
300 sec.
Ch5 span
300 sec.
Replacement
300 sec.
When auto-calibration is performed with the above settings:
:Auto calibration
in progress
Sunday
12
:
00
Interval
Monday
12
:
00
Tuesday
12
:
00
300 sec. 300 sec.
Ch4
Span cal.
Ch5
Span cal.
300 sec.
Replace
ment
Flow time
Zero cal.
350 sec.
Ch1
Span cal.
Ch2
Span cal.
350 sec. 350 sec. 350 sec.
Ch3
Span cal.
(This is an example when Ch1 to Ch5 are set to "Enable" in
Manual Zero Calibration Setting of 5.1.1 Calibration channel
setting. and the oxygen analyzer is a built-in oxygen analyzer.)
Figure 5.2
l
Setting range
Interval
1 to 99 hours or 1 to 40 days (default 7 days)
Flow time (zero gas, replacement time) 60 to 900 seconds (default 180 seconds)
Flow time (each span gas)
5 to 900 seconds (default: 180 seconds)
CAUTION
• When auto-calibration starts, the measurement screen is automatically displayed.
• After the power is turned off (including power failure, etc.) and then turned on again, the next
automatic calibration is performed at the time set for the start time, and then at intervals of
cycles.
• Regardless of whether auto-calibration is On or Off, auto-calibration can be performed from
the screen or by remote start.