IM 11G06A01-02EN
5-56
5. SCREENS
1st Edition : Oct. 27, 2023-00
5.6.4
Valve check (Auto zero cal.)
The operation of the valve at the time of auto zero calibration can be checked. Switch the state of
the valve and check the measured value at that time to confirm that there is no problem with the
valve operation. Before checking, configure the contact output setting and assign the necessary
functions to auto zero calibration. (For contact output setting, refer to "5.2.3 Contact Output
Setting.")
1 Tap icon on the measurement screen.
The menu screen appears.
2 Touch "Maintenance".
The maintenance menu will appear.
3 Select "Valve check (Auto zero cal.)".
The Valve check (Auto zero cal.) screen will appear
4 The combo box
shows the currently selected valve status.
Tap the combo box and select the valve status you wish to check.
The display switches to the selected valve state.
l
Description of valve selection
Meas. Valve status during measurement
Zero Valve status during zero validation