IM 11G06A01-02EN
5-27
5. SCREENS
1st Edition : Oct. 27, 2023-00
5.2.3
Contact output setting
The contact output setting is used to assign functions to the contact output terminals.
(1) Tap
on the Home/Main screen > Menu
(2) Tap [Setting] > Settings
(3) Tap [Contact output settings].
(4) Tap each item to set.
l
Items to set
Contact output x
Settings for functions to be assigned to contact output terminals
(x: 1 to 17)
l
Functions that can be assigned to contact outputs
See Table 5.1 for the functions that can be assigned to the contact outputs. Multiple functions can
be assigned to a single contact, or a single function can be assigned to multiple contacts.
To assign multiple functions to a single contact, select multiple functions in the function selection
dialog. When multiple functions are assigned, if any one of the functions is closed, that output
terminal will be closed. (OR Condition)
l
Limitations on contact output settings
For contact outputs 1 to 4, functions are fixedly assigned and cannot
be changed by the user. The assignments are as follows
Contact output 1
Equipment abnormality
Contact output 2
Calibration abnormality
Contact output 3
During auto-calibration
Contact output 4
For zero calibration