IM 11G06A01-02EN
5-4
5. SCREENS
1st Edition : Oct. 27, 2023-00
5.1.2
Auto Calibration Setting
IR800G/IR810
G
automatically calibrates the zero and span points at set times.
(1) Tap
on the Home/Main screen > Menu
(2) Tap [Cal/Alarm Setting] > Cal/Alarm setting
(3) Tap [Auto calibration setting] > Auto calibration setting
(4) Tap each item to set.
Figure 5.1
l
Items to set
On/Off
On/Off setting of the automatic calibration
First start day Set the starting day of the week for the first calibration
First start time [hhmm] Set what time and minute to start the first calibration
Interval
The time interval between the last calibration and the next one.
Interval unit
Set the unit of cycle (hours/day)
Gas Channel Gas flow path (analyzer/sampling system) setting for automatic calibration (gas
flow path for manual calibration also follows this setting)
Flow time
Time to replace with calibration gas and time to replace sample gas after
calibration is completed (set for each calibration gas; see Gas flow time setting.