<7. HART Communication>
7-22
IM 01C25W05-01EN
(2) Setting the burst mode
Recall a burst mode parameter, Burst Message 1, 2, or 3, and set the parameter according to the following flow
chart.
Burst
Command?
Burst Msg
Trigger Mode?
Update Period
Max Update Period
Burst Command
Burst mode
Burst Msg Trigger Mode
Continuous
Window
Rising
Falling
On-change
Send to transmitter
Cmd9, Cmd33
a. Burst Command
b. Burst Variables
c. Update period and Max
Update Period by “Set
Burst Period”
d. Burst Msg Trigger Mode
and Burst trigger level by
“Set Burst Trigger”
F0728.ai
e. Burst Mode
Cmd1, Cmd2,
Cmd3, Cmd48
Burst Variable Code
Burst Trigger Level
Send to transmitter
Send to transmitter
Send to transmitter
Send to transmitter