![YOKOGAWA DPharp EJXC40A User Manual Download Page 128](http://html1.mh-extra.com/html/yokogawa/dpharp-ejxc40a/dpharp-ejxc40a_user-manual_3660876128.webp)
<9. Parameter Summary>
9-1
IM 01C25W05-01EN
9. Parameter Summary
Table 9.1
Parameter Summary
Function
Label
Item
Description
Factory Default
Value
R/W
*1
Analog output AO Alrm typ
Output during hardware
failure
Indicates the output level during hardware
errors (including CPU errors) (Hi, Lo)
Depends on the
hardware switch
R
AO lower limit
Lower limit of analog
output
Sets the lower limit of output current.
(3.80 to 21.60 mA)
3.80 mA
W
AO upper limit
Upper limit of analog
output
Sets the upper limit of output current.
(3.80 to 21.60 mA)
21.60 mA
W
Auto recover
Auto recover from
hardware error
Specifies the action to take when the cause
of sensor error is removed. (On, Off)
On
W
Analog output
trim
Clear D/A trim
Output adjustment clear
Clears the analog output adjustment level.
M
D/A trim
D/A output adjustment
Performs analog output adjustment for 4 mA
and 20 mA.
M
Scaled D/A trim
Scaled D/A
output adjustment
Performs current output adjustment for 4 mA
and 20 mA. Scales and displays the output
value.
M
Channel flags
Analog channel flag
Indicates the analog channel flag.
0x00
R
Bi-direction
mode
Bi-dir mode
Bidirectional mode
Sets bidirectional flow mode on or off.
(On, Off)
Off
W
Burst
mode
Burst mode
Burst mode
Switches to or out of burst mode. (Wired
HART Enabled, Off)
Off
W
Burst Command
Burst Command
Cmd 1: PV
Cmd 2: % range/current
Cmd 3: Dyn vars/current
Cmd 9: Device vars w/ status
Cmd 33: Device Variables
Cmd 48: Read Additional Device Status.
Cmd 1: PV
W
Burst Variable
Code
Device variable of burst
message
Sets the device variable of burst message.
Up to 8 slots.
W
Set Burst Trigger
Burst trigger setting
Sets the burst trigger mode and trigger
value.
M
Set Burst Period
Burst period setting
Sets the burst period and longest burst
update time.
M
Burst Msg Trigger
Mode
Burst message trigger
mode
Indicates the burst message trigger mode.
(Continuous, Window, Rising, Falling, On-
change)
Continuous
R
Burst Trigger Level Trigger value
Indicates the trigger value.
0.000000
R
Update Period
Update period for Burst
mode
Indicates the shortest burst update time.
Burst Message
1: 0.500 s
2: 60.000 s
3: 60.000 s
R
Max Update
Period
Longest update period for
Burst mode
Indicates the longest burst update time.
3600.000 s
R
*1:
R:read only, W:read and write, M:method, G:applicable for option code /DG6.