<9. Parameter Summary>
9-9
IM 01C25W05-01EN
Function
Label
Item
Description
Factory Default
Value
R/W
*1
Status
Device status
Field device status
Indicates the status.
R
Diagnostic status 0 Diagnostic status
Indicates device diagnostic status
R
Diagnostic status 1 Diagnostic status
Indicates device diagnostic status
R
Status group 0
Hardware status
information
Indicates module 1 side hardware failure
errors.
R
Status group 1
Hardware status
information
Indicates module 1 side hardware failure
errors.
R
Status group 2
Process status
information
Indicates module 1 side process errors
(outside-the-specification-range error)
R
Status group 3
Process status
information
Indicates module 1 side process errors
(outside-the-specification-range error)
R
Status group 4
Process status
information
Indicates module 1 side process errors
(outside-the-measuring-range error)
R
Status group 5
Configuration status
information
Indicates module 1 side configuration errors.
R
Status group 14
Configuration status
information
Indicates module 1 side configuration errors.
R
Status group 15
Process status
information
Indicates diagnostic alarms.
R
Status group 16
Process status
information
Indicates diagnostic alarms.
R
Status group 17
Process status
information
Indicates the simulation status.
R
Status group 18
Hardware status
information
Indicates module 2 side hardware failure
errors.
R
Status group 19
Process status
information
Indicates module 2 side process errors
(outside-the-specification-range error)
R
Status group 20
Process status
information
Indicates module 2 side process errors
(outside-the-specification-range error)
R
Status group 21
Process status
information
Indicates module 2 side process errors
(outside-the-measuring-range error)
R
Status group 22
Configuration status
information
Indicates module 2 side configuration errors.
R
Status group 23
Configuration status
information
Indicates module 2 side configuration errors.
R
Ext dev status
Extended field device
status
Indicates the status.
R
Cfg chng count
Configuration change
counter
Indicates the number of times parameter
configuration has changed.
0
R
Reset Cfg chng
flag
Reset configuration
change flag
Resets the configuration change flag.
M
Current Date
Data timestamp
Indicates the UTC year, month, and day
when data was read. The reference date is
set from a host system.
1970/01/01
R
Current Time
Data timestamp
Indicates the UTC time when data was
read. The reference date is set from a host
system.
00:00:00
R
Tag number
Tag
Tag number
Indicates or sets the tag number.
Up to 8 alphanumeric characters
As specified in
order
W
Long tag
Long tag number
Indicates or sets the long tag number.
Up to 32 alphanumeric characters
As specified in
order
W
*1: