<9. Parameter Summary>
9-6
IM 01C25W05-01EN
Function
Label
Item
Description
Factory Default
Value
R/W
*1
Device
variable status
DP Limit Status
Device variable limit
status
Indicates the limit status of the differential
pressure. (Table 8.7)
R
DP Data Quality
Device variable data
quality
Indicates the quality status of the differential
pressure. (Table 8.7)
R
Pres 1 Limit Status Device variable limit
status
Indicates the limit status of the module 1 side
pressure.
R
Pres 1 Data
Quality
Device variable data
quality
Indicates the quality status of the module 1
side pressure.
R
Pres 2 Limit Status Device variable limit
status
Indicates the limit status of the module 2 side
pressure.
R
Pres 2 Data
Quality
Device variable data
quality
Indicates the quality status of the module 2
side pressure.
R
Temp 1 Limit
Status
Device variable limit
status
Indicates the limit status of the module 1 side
capsule temperature.
R
Temp 1 Data
Quality
Device variable data
quality
Indicates the quality status of the module 1
side capsule temperature.
R
Temp 2 Limit
Status
Device variable limit
status
Indicates the limit status of the module 2 side
capsule temperature.
R
Temp 2 Data
Quality
Device variable data
quality
Indicates the quality status of the module 2
side capsule temperature.
R
PV % rnge Limit
Status
Device variable limit
status
Indicates the limit status of the percentage
value.
R
PV % rnge Data
Quality
Device variable data
quality
Indicates the quality status of the percentage
value.
R
Loop current Limit
Status
Device variable limit
status
Indicates the limit status of the current (mA).
R
Loop current Data
Quality
Device variable data
quality
Indicates the quality status of the current
(mA).
R
DP update time
period
Differential pressure
update period
Indicates the differential pressure update
period.
90 ms
R
Pres update time
period
Pressure update period
Indicates the pressure update period.
90 ms
R
Temp update time
period
Temperature update
period
Indicates the temperature update period.
1s
R
Self-
diagnostics
Self test
Self-diagnostics
Executes self-diagnostics.
M
Master test
Master test
Executes a self-test after performing a
software reset.
M
*1: