<8. Maintenance>
8-22
IM 01C25W05-01EN
Integral
Indicator
Display
Display on
Configuration
Tool
Device
Status
Extended
Device
Status
NE107
(Default
Value)
Value and Status (Data Quality, Limit Status)
Differential
pressure
Module 1
Module 2
%
Range
Current
Output
Pressure
Capsule
Temperature Pressure
Capsule
Temperature
AL.57
P2. ADJ
Pres 2 Zero
Trim Err
(AL-57)
—
Function Check
(0x20)
C
Value:
Measured
value
Status: Poor
Accuracy,
Not Limited
Value: Measured value
Status: Good, Not
Limited
Value:
Measured
value
Status:
Poor
Accuracy,
Not Limited
Value:
Measured
value
Status: Good,
Not Limited
[In the case of PV]
Value: Measured
value
Status: Poor
Accuracy, Not
Limited
[In the case of
other than PV]
Value: Measured
value
Status: Good, Not
Limited
Pres 2 Span
Trim Err
(AL-57)
C
—
Pres 2 Fixed
Mode
[In the case of
PV]
Loop Current
Fixed (0x08)
C
Value: Measured value
Status: Good, Not Limited
Value: Measured value
Status: Good , Constant
[In the case of PV]
Value: Hold value
Status: Good,
Constant
[In the case of
other than PV]
Value: Measured
value
Status: Good, Not
Limited
AL.60
SC. CFG
S.C. Config
Error
(AL-60)
—
Function Check
(0x20)
C
Value: Measured value
Status: Good, Not Limited
Value: Measured
value
Status: Bad, Not
Limited
AL.79
OV.DISP
—
—
—
N
Value: Measured value
Status: Good, Not Limited
*1: Depends on the setting of hardware switch
*2: Depends on the direction of rangeover (high or low)