<9. Parameter Summary>
9-10
IM 01C25W05-01EN
Function
Label
Item
Description
Factory Default
Value
R/W
*1
Temperature
compensation
T.Z. Cmp mode 1
Module 1 side
temperature zero
compensation mode
Selects the module 1 side temperature zero
compensation mode (On, Off)
Off
W
T.Z. Cmp mode 2
Module 2 side
temperature zero
compensation mode
Selects the module 2 side temperature zero
compensation mode (On, Off)
Off
W
Temp 1 zero
Module 1 side
temperature zero
compensation coefficient
Sets the temperature gradient of the module
1 side temperature zero compensation
coefficient. (–99.999 to 99.999%/°C)
0.000%/degC
W
Temp 2 zero
Module 2 side
temperature zero
compensation coefficient
Sets the temperature gradient of the module
2 side temperature zero compensation
coefficient. (–99.999 to 99.999%/°C)
0.000%/degC
W
Temperature
sensor
Amp temp
Module 1 side amplifier
temperature
Indicates the amplifier (amplifier assembly)
temperature.
R
Temp 2 fixed val
Module 2 side capsule
temperature fixed value
Sets the module 2 side capsule temperature
fixed value.
23 degC
W
Unit
DP unit
Differential pressure unit Sets the differential pressure unit.
W
Pres 1 unit
Module 1 side pressure
unit
Sets the module 1 side pressure unit.
W
Pres 2 unit
Module 2 side pressure
unit
Sets the module 2 side pressure unit.
W
Temp unit
Temperature unit
Sets the temperature unit.
W
Write
protection
Enable wrt 10min
Write protection release
Temporarily releases write protection
(10 minutes). (Enter an 8 alphanumeric
character parameter.)
M
New password
Password setting
Sets the write protection release password
and enables or disables the function.
(8 alphanumeric characters)
M
Write protect
Communication writing
enabled/disabled
Indicates the write protection status. (Yes,
No)
No
R
Device
variable
Loop current
Output current
Indicates the 4 to 20 mA output current.
R
PV % rnge
Percentage output
Indicates the 4 to 20 mA output as a
percentage.
R
DP
Differential pressure
Indicates the differential pressure.
Pres 1
Module 1 side pressure
Indicates the module 1 side pressure.
R
Pres 2
Module 2 side pressure
Indicates the module 2 side pressure.
R
Temp 1
Module 1 side capsule
temperature
Indicates the temperature at the module 1
side capsule temperature sensor.
R
Temp 2
Module 2 side capsule
temperature
Indicates the temperature at the module 2
side capsule temperature sensor.
R
Device
variable
simulation
Device vars
simulate
Device variable
simulation execution
Execute the simulation.
M
*1:
R:read only, W:read and write, M:method, G:applicable for option code /DG6.