<8. Maintenance>
8-18
IM 01C25W05-01EN
Integral
Indicator
Display
Display on
Configuration
Tool
Device
Status
Extended
Device
Status
NE107
(Default
Value)
Value and Status (Data Quality, Limit Status)
Differential
pressure
Module 1
Module 2
%
Range
Current
Output
Pressure
Capsule
Temperature Pressure
Capsule
Temperature
AL.03
CAP2.
ER
Pres 2 Sensor
Failure
(AL-03)
Device
Malfunction
(0x80)
Failure
(0x08)
F
Value: Hold
value
Status: Bad,
Constant
Value: Measured value
Status: Good, Not
Limited
Value:
Hold value
Status:
Bad,
Constant
Value:
Measured
value
Status: Good,
Not Limited
Value: Hold value
Status: Bad |
Low Limited/High
Limited
*1
Temp 2 Sensor
Failure
(AL-03)
F
Value:
Measured
value
Status: Poor
Accuracy,
Not Limited
Value:
Measured
value
Status:
Poor
Accuracy,
Not Limited
Value:
Measured
value
Status: Bad,
Not Limited
Cap 2
EEPROM
Failure
(AL-03)
F
Value: Hold
value
Status: Bad,
Constant
Value:
Hold value
Status:
Bad,
Constant
Value:
Measured
value
Status: Bad,
Not Limited
AL.04
AMP2.
ER
Amp 2
EEPROM
Failure
(AL-04)
F
Value:
Measured
value
Status: Bad,
Not Limited
CPU 2 Board
Failure
(AL-04)
F
Value: Hold
value
Status: Bad,
Constant/
Value:
Measured
value
Status: Bad,
Not Limited
AL.05
MDL2.
ER
Module 2
Failure
(AL-05)
F
Value: Hold
value
Status: Bad,
Constant
AL.06
COMM.
ER
Intercom
Failure
(AL-06)
F
—
No Device ID —
F
Value: Measured value
Status: Good, Not Limited
*1: Depends on the setting of hardware switch
*2: Depends on the direction of rangeover (high or low)