<8. Maintenance>
8-24
IM 01C25W05-01EN
Integral
Indicator
Display
Display on
Configuration
Tool
Device
Status
Extended
Device
Status
NE107
(Default
Value)
Value and Status (Data Quality, Limit Status)
Differential
pressure
Module 1
Module 2
%
Range
Current
Output
Pressure
Capsule
Temperature Pressure
Capsule
Temperature
AL.91
DP. SIM
DP Simulate
Mode
(AL-91)
[In the case of
PV]
Loop Current
Fixed (0x08)
Function Check
(0x20)
C
Value and
Status:
Simulated
value
Value: Measured value
Status: Good, Not Limited
[In the case of PV]
same as PV Value
and Status
[In the case of
other than PV]
Value: Measured
value
Status: Good, Not
Limited
AL.91
P1. SIM
Pres 1
Simulate Mode
(AL-91)
C
Value:
Measured
value
Status:
Good, Not
Limited
Value and
Status:
Simulated
value
Value: Measured value
Status: Good, Not Limited
AL.91
P2. SIM
Pres 2
Simulate Mode
(AL-91)
C
Value: Measured value
Status: Good, Not Limited
Value and
Status:
Simulated
value
Value:
Measured
value
Status: Good,
Not Limited
AL.91
T1. SIM
Temp 1
Simulate Mode
(AL-91)
—
C
Value: Measured value
Status: Good, Not
Limited
Value and
Status:
Simulated
value
Value: Measured value
Status: Good, Not Limited
AL.91
T2. SIM
Temp 2
Simulate Mode
(AL-91)
C
Value: Measured value
Status: Good, Not Limited
Value and
Status:
Simulated
value
Value: Measured
value
Status: Good, Not
Limited
AL.91
PCT.SIM
PV % rnge
Sim Mode
(AL-91)
Loop Current
Fixed
(0x08)
C
Value: Measured value
Status: Good, Not Limited
Value and Status:
Simulated value
AL.91
AO. SIM
Loop current
Sim Mode
(AL-91)
C
Value: Measured value
Status: Good, Not Limited
Value and Status:
Simulated value
AL.92
STS.SIM
Status Sim
Mode
(AL-92)
—
—
N
Value: Measured value
Status: Good, Not Limited
*1: Depends on the setting of hardware switch
*2: Depends on the direction of rangeover (high or low)