<9. Parameter Summary>
9-11
IM 01C25W05-01EN
Function
Label
Item
Description
Factory Default
Value
R/W
*1
Process
variables
Change PV Assgn PV assignment
Specifies the variable to assign to PV
(primary variable). (DP, Pres 1, Pres 2)
M
PV is
PV assignment
Indicates the variable currently assigned to
PV.
DP
R
Change SV Assgn SV assignment
Specifies the variable to assign to
SV(secondary variable).(DP, Pres 1, Pres 2,
Temp 1, Temp 2)
M
SV is
SV assignment
Indicates the variable currently assigned to
SV.
Pres 1
R
Change TV Assgn TV assignment
Specifies the variable to assign to
TV(Teritary variable).(DP, Pres 1, Pres 2,
Temp 1, Temp 2)
M
TV is
TV assignment
Indicates the variable currently assigned to
TV.
Pres 2
R
Change QV Assgn QV assignment
Specifies the variable to assign to
QV(Quaternary variable).(DP, Pres 1, Pres
2, Temp 1, Temp 2)
M
QV is
QV assignment
Indicates the variable currently assigned to
QV.
Temp 1
R
*1:
R:read only, W:read and write, M:method, G:applicable for option code /DG6.