IM DL850E-01EN
Display Type (Display Type)
When you measure voltage using a voltage module or measure strain using a strain module, you can display the
linearly scaled results using one of the following two methods.
Exponent: Values are displayed in exponential notation.
Floating: Values are displayed as decimal numbers.
For Decimal Number, set the number of digits after the decimal point to Auto or a number from 0 to 3. For
SubUnit, set the unit prefix to Auto, p, n, μ, m, None, k, M, G, or T.
• If you set Decimal Number to a number from 0 to 3, the specified number of digits after the decimal point
are displayed. If you select Auto, all numbers will be displayed using five digits (for example: 1.0000,
250.00). The default setting is Auto.
• If you set SubUnit to a setting other than Auto, numbers will be displayed with the specified unit prefix.
If you select Auto, the DL850E/DL850EV will automatically select appropriate unit prefixes. The default
setting is Auto. The DL850E/DL850EV displays values in exponential notation if it is unable to display them
as decimal numbers.
Scale Value Display
You can display the linearly scaled values of the upper and lower vertical limits of each channel.
You can turn the scale value display on and off in the DISPLAY menu.
• Linear scaling cannot be performed for measurements of temperature or acceleration.
• The following waveforms cannot be linearly scaled.
Accumulated waveforms (except for the most recent)
• You can configure linear scaling for each channel.
• The specified scaling coefficient A and offset B are retained even after linear scaling is set to OFF.
• Computations are performed on the linearly scaled values.
2 Vertical Axis