2-37
IM DL850E-01EN
The following 9 setup items are available.
Voltage Range (V Range)
You can set the input voltage range (±FS) to one of the options below.
• When the probe attenuation (Probe) is set to 1:1
±1 V, ±2 V, ±5 V, ±10 V, ±20 V, ±50 V (±FS)
• When the probe attenuation (Probe) is set to 10:1
±10 V, ±20 V, ±50 V, ±100 V, ±200 V, ±500 V (±FS)
Input Coupling (Coupling)
You can set the input coupling to DC or AC.
Probe Type (Probe)
You can set the probe type to 1:1 or 1:10.
Bandwidth (Bandwidth)
You can set the bandwidth limit to 100 Hz, 1 kHz, 10 kHz, 100 kHz, or Full. You cannot select Full when Preset is
set to AC100V or AC200V.
Threshold Level (Threshold)
You can set the level within the FS of the input voltage range. The resolution is the value that corresponds to 1%
of the FS.
Hysteresis (Hysteresis)
You can set the hysteresis to ±1%, ±2.5%, or ±5% of the FS of the input voltage range.
Slope (Slope)
Select (rising) or (falling).
Chattering Elimination (Chatter Elimination)
Eliminates the chattering that occurs in cases such as when the contact input is turned on or off. The changes in
the signal over the specified time can be discarded. You can set the interval to a value from 0 to 1000 ms (in 1
ms steps). When 0 ms is selected, chattering elimination is turned off.
Chattering elimination works on both rising and falling edges.