9-1
IM DL850E-01EN
9 Cursor Measurement
You can move cursors on the waveforms displayed on the screen to view the measured values at the points
where the cursors intersect the waveforms. You can select whether to measure the P-P compressed data values
on the screen or the data values that have been acquired in the acquisition memory.
Window Selection (Select Window)
Select the window to perform cursor measurement in. This option appears when X-Y or FFT waveforms are
being displayed.
•
: Displays T-Y waveform cursor-measurement values.
•
: Displays the cursor-measurement values for the waveform in an X-Y window.
•
: Displays the cursor-measurement values for the waveform in an FFT window.
T-Y Waveforms
Cursor Types (Type)
The following types of T-Y waveform cursors are available.
• OFF: Cursor measurement is not performed.
•
Horizontal cursors (Horizontal)
: Two horizontal cursors are used to measure vertical values.
•
: Two vertical cursors are used to measure time values.
•
Horizontal and vertical cursors (H & V)
: Two horizontal cursors and two vertical cursors are used to measure
vertical and time values.
•
: Four marker cursors that move on the waveform are used to measure waveform
values.
•
: Two angle cursors are used to measure angles.
Horizontal Cursors (Horizontal) - T-Y waveforms
Two dashed lines (horizontal cursors) appear on the horizontal axis. You can measure the vertical value at the
position of each horizontal cursor and measure the level difference between the horizontal cursors.
Cursor1
Cursor2
Δ
Y
Y1
Y2
Measurement Source Waveform (Trace)
Set the measurement source waveform to one of the waveforms below.
CH1 to CH16,
1
16chVOLT,
2
16chTEMP/VOLT,
2
CAN,
3
LIN,
3
Math1 to Math8
1 You can select the channel of an installed module. However, you cannot select the channel of a logic module.
2 When a 16-CH voltage input module or 16-CH temperature/voltage input module is installed. After you select
16chVOLT or 16chTEMP/VOLT, select a sub channel.
3 On the DL850EV when a CAN bus monitor module or CAN & LIN bus monitor module is installed. After you
select CAN or LIN, select a sub channel. You cannot select waveforms whose data type (Value Type) is set to
Logic.