13-4
IM DL850E-01EN
Waveform Parameters (Parameter)
Set the upper and lower limits for automated measurement values of waveform parameters, and perform GO/
NO-GO determination based on whether the values are within or outside of the limits.
Judgment Conditions (Judgement Setup)
For each of 16 judgment conditions, you can set the source waveform, waveform parameter, and upper
and lower waveform parameter limits. You can also set the judgment logic, action condition, sequence, and
acquisition count and enable or disable synchronization with an external start signal.
Judgment Criterion (Mode)
Select the judgment criterion from the following options.
• X: The condition is not used for GO/NO-GO determination.
• IN: The DL850E/DL850EV returns a GO result when the parameter is within the specified limits. The DL850E/
DL850EV returns a NO-GO result when the parameter is outside the specified limits.
• OUT: The DL850E/DL850EV returns a GO result when the parameter is outside the specified limits. The
DL850E/DL850EV returns a NO-GO result when the parameter is inside the specified limits.
Source Waveform (Trace)
Set the waveform to use for GO/NO-GO determination to one of the waveforms below.
CH1 to CH16
1
, Math1 to Math8
1 You can select the channel of an installed module. However, you cannot select the channel of a 16-CH
voltage input module, 16-CH temperature/voltage input module, CAN bus monitor module, or CAN & LIN bus
monitor module.
Waveform Parameter (Item)
You can use all automatically measured waveform parameters as reference conditions. You can perform GO/
NO-GO determination on up to 16 parameters at the same time.
Upper and Lower Parameter Limit Settings (Upper/Lower)
Selectable range: −9.9999E+30 to 9.9999E+30.
Judgment Logic (Logic), Action Condition (ActCondition), Sequence (Sequence), Acquisition Count
(Acquisition Count), and External Start (Ext Start)
Action (Action)
Determination Period (Time Range1 and Time Range2)
GO/NO-GO Determination Results
The results of GO/NO-GO determination (number of determinations, number of failures) appear at the bottom of
the screen.
Determination
count
Failure count
Exe Count: 10
Fail Count: 4
Result: XXXXOOOO--------
Measured values for
each parameter
In this example, reference conditions 1 and 4 are not met while reference conditions 5
to 8 are met.
Indication of whether reference
conditions 1 to 16 are met
XXXXOOOO--------
1,2,3, ...,16: Reference conditions
O: Condition met
X: Condition not met
- : No condition specified
13 GO/NO-GO Determination