9-6
IM DL850E-01EN
Angle Cursors (Degree) - T-Y waveforms
You can measure time values and convert them to angles. On the time axis, set the zero point (Ref Cursor1
position), which will be the measurement reference, the end point (Ref Cursor2 position), and the reference
angle that you want to assign to the difference between Ref Cursor1 and Ref Cursor2. Based on this reference
angle, you can measure the angle between two angle cursors (Cursor1 and Cursor2).
Cursor1
Cursor2
Ref Cursor1
Ref Cursor2
Reference angle
Measurement Source Waveform (Trace)
Set the measurement source waveform to one of the waveforms below.
CH1 to CH16,
1
16chVOLT,
2
16chTEMP/VOLT,
2
CAN,
3
LIN,
3
Math1 to Math8, All
4
1 You can select the channel of an installed module.
2 When a 16-CH voltage input module or 16-CH temperature/voltage input module is installed. After you select
16chVOLT or 16chTEMP/VOLT, select a sub channel.
3 On the DL850EV when a CAN bus monitor module or CAN & LIN bus monitor module is installed. After you
select CAN or LIN, select a sub channel.
4 Cursor measurement is performed on all channels.
Moving the Cursors (Cursor1/Cursor2)
Use Cursor1 and Cursor2 to move the cursors.
Taking the center of the waveform display window to be 0 div, you can move the cursors within the range of −5 to
5 div in steps of the following size: Time/div × 10 ÷ display record length.
Cursor Jumping (Cursor Jump)
You can make Cursor1 and Cursor2 jump to the center of the specified zoom window.
Cursor1 to Zoom1: Cursor1 jumps to the Zoom1 window.
Cursor1 to Zoom2: Cursor1 jumps to the Zoom2 window.
Cursor2 to Zoom1: Cursor2 jumps to the Zoom1 window.
Cursor2 to Zoom2: Cursor2 jumps to the Zoom2 window.
Reference Angle (Ref Value)
Set the reference angle you want to assign to the range defined by Ref Cursor1 and Ref Cursor2.
Selectable range: 1 to 720
References (Ref1/Ref2)
Set the zero point (Ref Cursor1) and the end point (Ref Cursor2).
Taking the center of the waveform display window to be 0 div, you can move the cursors within the range of −5 to
5 div in steps of the following size: Time/div × 10 ÷ display record length.
9 Cursor Measurement