13-2
IM DL850E-01EN
Zone Settings
When Edit is set to Whole, you can set the upper, lower, left, and right boundaries of the waveform zone. When
Edit is set to Part, you can set the upper and lower boundaries of the zone.
• Upper and lower boundaries (Upper and Lower)
Selectable range: ±10 div vertically from the base waveform
When Edit is set to Part, you can set the upper and lower boundaries of the area between Time Range1 and
Time Range2.
• Left and right sides (Left and Right): These settings can be configured only when Edit is set to Whole.
Selectable range: ±5 div from the center of the screen
• Time range 1 and time range 2 (Time Range1 and Time Range2): These settings can be configured only
when Edit is set to Part.
Selectable range: ±5 div on the time axis
Save Destination (Store as)
You can select one of the following save destinations for the waveform zone.
• Zone 1 to zone 6 (Zone1 to Zone6): The save destination is changed to the selected zone number.
• Cancel (Cancel): The save destination is not changed.
Saving a Waveform Zone (Execute Store)
Save the waveform zone.
Ending Editing (Quit)
Exit the waveform zone editor. If you do not press the Execute Store soft key to save the zone that you have
edited, it will be lost.
Judgment Conditions (Judgement Setup)
For each of 16 judgment conditions, you can set the source waveform, zone number, and judgment criterion.
You can also set the judgment logic, action condition, sequence, and acquisition count and enable or disable
synchronization with a remote signal.
Judgment Criterion (Mode)
Select the judgment criterion from the following options.
• X: The condition is not used for GO/NO-GO determination.
• IN: The DL850E/DL850EV returns a GO result when the source waveform is within the GO/NO-GO
determination zone. If even part of the source waveform is outside of the determination zone, the DL850E/
DL850EV returns a NO-GO result.
• OUT: The DL850E/DL850EV returns a GO result when the entire source waveform is outside the GO/NO-
GO determination zone. If even part of the source waveform is inside the determination zone, the DL850E/
DL850EV returns a NO-GO result.
Source Waveform (Trace)
Set the waveform to use for GO/NO-GO determination to one of the waveforms below.
CH1 to CH16
1
, Math1 to Math8
1 You can select the channel of an installed module. However, you cannot select the channel of a logic module,
16-CH voltage input module, 16-CH temperature/voltage input module, CAN bus monitor module, or CAN &
LIN bus monitor module.
Zone Number (Zone No.)
Select the number of the waveform zone you want to use for GO/NO-GO determination from the range indicated
below.
Zone 1 to zone 6 (Zone1 to Zone6)
Determination Logic (Logic)
You can select the determination logic from the following options.
• AND: The actions are performed when all the conditions from 1 to 16 are met.
• OR: The actions are performed when a condition from 1 to 16 is met.