2-9
IM DL850E-01EN
Zoom Method (V Scale)
You can choose the method for zooming the waveform vertically.
• DIV: The waveform is zoomed by a set magnification.
• SPAN: The waveform is zoomed to fit within specified upper and lower display limits.
Zooming by Setting a Magnification (V Zoom)
When V Scale is set to DIV, you can set the vertical magnification and enlarge or reduce the waveform along the
vertical axis. This method is useful when you want to change the vertical scale of the waveform after it has been
displayed.
Zoom Factor (V Zoom)
The magnifications that you can choose from are listed below.
x0.1, x0.111, x0.125, x0.143, x0.167, x0.2, x0.25, x0.33, x0.4, x0.5, x0.556, x0.625, x0.667, x0.714, x0.8, x0.833,
x1, x1.11, x1.25, x1.33, x1.43, x1.67, x2, x2.22, x2.5, x3.33, x4, x5, x6.67, x8, x10, x12.5, x16.7, x20, x25, x40,
x50, x100
The range of zoom magnifications that you can choose varies under special circumstances as described below.
When using the 701267 (HV (with RMS))
50 V/div: x0.25 to x100
100 V/div: x0.5 to x100
200 V/div: x1 to x100
When measuring acceleration using the 701275 (ACCL/VOLT)
x0.5 to x50
When using the 701280 (FREQ)
x0.33 to x100
When using the 720230 (LOGIC)
x0.1 to x3.33
Zoom Position
Zooming is centered on the vertical position.
When V Zoom is set to ×2
Vertical position mark
1 div = 0.5 V
Ground level mark
1 div = 1 V
2 Vertical Axis