2-20
IM DL850E-01EN
Temperature Measurement
For temperature measurement, the items that have to be set for each input signal (CH1 to CH16) include vertical
scales, the vertical positions, input coupling, thermocouple type, the bandwidth limit, trace settings, temperature
settings, and the temperature unit.
You can connect a thermocouple to one of the temperature modules, 701261 (UNIVERSAL), 701262 (UNIVERSAL
(AAF)), or 701265 (TEMP/HPV), and measure temperatures.
For information about how to connect thermocouples, see section 3.8 in the
Getting Started Guide
, IM DL850E-
03EN.
Waveform Display (Display)
Select whether to display each channel’s input signal waveform.
• ON: Displays the waveform
• OFF: Does not display the waveform
Labels (Label)
Input Coupling (Coupling)
Set the coupling to TC when you are measuring temperature using the 701261 (UNIVERSAL), 701262
(UNIVERSAL (AAF)), or 701265 (TEMP/HPV).
To measure voltage, set the input coupling to an appropriate voltage measurement setting.
Thermocouple Type (Type)
Select the appropriate thermocouple type for thermocouple that you are using.
You can select thermocouple type from the options listed below.
When the input terminals are open, the displayed value is less than or equal to the lower limit of the
measurement range.
Type
Measurement Range
K
−
200 to 1300°C
E
−
200 to 800°C
J
−
200 to 1100°C
T
−
200 to 400°C
L
−
200 to 900°C
U
−
200 to 400°C
R
0 to 1700°C
S
0 to 1700°C
B
400 to 1800°C
N
0 to 1300°C
W
0 to 2300°C
Au7Fe (AuFe-chromel)
0 to 300 K
Bandwidth (Bandwidth)
You can set the bandwidth limit to 2 Hz, 8 Hz, 30 Hz, or Full.
2 Vertical Axis