IM DL850E-01EN
Real Time Math (RealTime Math)
Turning Real Time Math On and Off
Select whether to use real time math.
• ON: Select this item to display a menu for configuring real time math. At the same time, real time math
execution begins.
• OFF: Select this item to display a menu for configuring the standard model. Real time math is not executed.
You can perform real time math operations in which the waveforms of input channels or the results of other real
time math operations are used as the math source waveforms. This is one of the features of the /G3 option. The
other feature is the
• Configure the settings for each channel. You can perform math operations on up to 16 channels at the same
time.
• When you turn real time math on, the real time math results are output to the real time math channels (the
channels that you have turned math on for). The waveforms of input channels that you have turned math on
for are not displayed. They are also not saved. For example, if you turn real time math on for input channel
CH2, CH2 becomes the RMath2 real time math channel, and the math results are displayed on the screen.
The data that is saved is that of the math result. If you want to display and save the waveform of the input
channel, set the real time math to a channel that has no input.
• You can set the real time math source waveforms to the input channels of 16-CH voltage input, 16-CH
temperature/voltage input, CAN bus monitor, and CAN & LIN bus monitor module, but you cannot set the
channel that the real time math result is output on to an input channel of a 16-CH voltage input, 16-CH
temperature/voltage input, CAN bus monitor, or CAN & LIN bus monitor module (there is no menu for turning
real time math on).
• For details on the modules whose channels you can set as real time math sources, see “Notes Regarding
Using the Digital Filter and Real Time Math”
• Even during waveform acquisition, you can set various math conditions, such as the operator or function (the
operation definition), the source waveforms, and the coefficients. However, if you change the conditions, the
measurement count (waveform acquisition count) is reset. The measurement count is displayed in the lower
left of the screen.
• By setting the waveform that results from math operations as a trigger source, you can trigger the DL850E/
DL850EV on the results.
• For details on the math expressions, delay, and settings, see the appendix in the
Real Time Math (/G3 option)
User’s Manual
, IM DL850E-51EN.
Labels (Label)
2 Vertical Axis