4-2
IM AQ1210-01EN
• Switching the Display between TRACE Mode and MAP Mode
3.
Tap the
TRACE
/
MAP
button to set the data display screen to TRACE mode.
When the data display screen is in TRACE mode, the button is MAP.
Executing an Averaged Measurement
4.
Tap
Wavelength
to set the wavelength.
5.
Press
AVG
. The measurement begins, and the data display screen shows the waveform.
During averaged measurement, the duration to average over is displayed in the top area of the screen.
Also, during measurement, a mark appears at the top of the display to indicate that the laser light is on.
When averaged measurements are completed, measurement is stopped automatically, event analysis is
executed, and the analysis results are displayed on the screen. If you press AVG again during averaged
measurement, measurement is stopped.
Note
• You can select the initial screen that appears when the instrument starts in system settings. For details on
operation, see the explanation in section 10.1.
• If distance range and pulse width are set to auto, the instrument automatically selects the optimal values
internally immediately before starting a measurement. Change these values if necessary. For details, see
section 2.1.
Soft Key Menu (during averaged measurement)
6.
Tap the soft key menu display button.
The menu that can be used during averaged measurement or the menu that can be used not during
averaged measurement is displayed automatically depending on the usage condition.
4.1 Measuring in TRACE Mode