6-9
IM AQ1210-01EN
Analyzing W
aveforms
1
2
3
4
5
6
7
8
9
10
11
App
Index
6 Point Markers
7.
Tap the
6 Point Markers
soft key. A soft key menu for the 6 markers method appears.
1
Marker
2
Marker
1
Sets marker
Places marker at the
cursor position
1
Sets marker
Places marker at the
cursor position
2
2
Return loss
Splice loss
Set marker Y2.
Places marker Y2 at the
cursor position
Sets marker
Places marker at the
cursor position
3
3
Marker Y2
3
Marker
See “2 Point Markers
(when the marker
mode is Marker)”
Marker Y1
Marker Y3
Set marker Y1.
Places marker Y1 at the
cursor position
Set marker Y3.
Places marker Y3 at the
cursor position
Note
• Set the markers in the order shown above, starting with the measurement start point (the distance
reference).
• The measured loss value differs depending on the specified approximation method.
• Set the correct position for
2
. The splice loss changes greatly depending on the position of
2
.
6.1 Operating Cursors and Markers