6.7 Combining Two Traces
Procedure
2-WayTrace Screen
1.
Use the
rotary knob
and
ENTER
to select
OTDR
.
2
.
Press the
Advanced Analysis
soft key and then the
2-WayTrace
soft key. The following screen
appears.
When the suffix code is -HE, press the
Analysis
soft key, the
Advanced Analysis
soft key and then
the 2-WayTrace
soft key.
Selects the current trace
(Trace 1, Trace 2)
Executes 2 way trace analysis
The event screen of 2 way trace analysis appears.
Load traces to display in Trace 1 and Trace 2.
The file operation screen appears
(see section 17.2).
Screen
(Trace+List, List)
See section 7.1.
Display type
(Reverse, Normal)
Selects the rotary knob
function (Cursor, Event)
Explanation
Loading Trace Data
The type of file that you can load is SOR.
Note
Traces that meet the following conditions can be combined.
• Event lists are available.
• The wavelengths and pulse widths of the two traces are the same.
• The distances from the measurement reference (S) to the end event (E) are the same.
Current Trace
The current trace can be set to one of the combined traces.
Cursors are displayed on the current trace.