Use this screen to specify the folder where files will be saved to.
Folder path of the folder that you want to save to
The number of folders and files in
the folder that you want to save to
Explanation
Measurement Wavelength
There are three instrument types, with the measurement light wavelengths listed below. Select a
wavelength from the available settings on the Instrument that you are using.
Model
Measurement Light Wavelength
Output Port
AQ1200A
SM 1310 nm, SM 1550 nm
Optical port 2
AQ1200B
SM 1625 nm
AQ1200C
SM 1650 nm
AQ1200E
SM 1310 nm, SM 1550 nm
Optical port 2
SM 1625 nm
Optical port 3
AQ1205A
SM 1310 nm, SM 1550 nm
Optical port 2
AQ1205E
SM 1310 nm, SM 1550 nm
Optical port 2
SM 1625 nm
Optical port 3
AQ1205F
SM 1310 nm, SM 1550 nm
Optical port 2
SM 1650 nm
Optical port 3
On the AQ1200A, AQ1200B, AQ1200C, and AQ1205A, the light for a single mode (SM) optical fiber is
emitted from optical port 2.
On the AQ1200E, AQ1205E, and AQ1205F, light with a 1310 nm or 1550 nm wavelength is transmitted
from optical port 2, and light with a 1625 nm or 1650 nm wavelength is transmitted from optical port 3.
Firmly connect the optical fiber to the port that the light with the selected wavelength will be transmitted
from.
Setting Wavelengths in Multi WL Measurement Mode
If the measurement mode is set to Multi WL, you can measure with both the 1310 nm wavelength
and 1550 nm wavelength at the same time.
Depending on the model, you can set the wavelengths as shown below.
Model
Wavelength 1
Wavelength 2
Wavelength 3
AQ1200A
SM 1310nm
SM 1550nm/OFF
─
SM 1550nm
SM 1310nm/OFF
─
AQ1200E
SM 1310nm
SM 1550nm/OFF
OFF
SM 1550nm
SM 1310nm/OFF
OFF
SM 1625nm
OFF
─
AQ1205A
SM 1310nm
SM 1550nm/OFF
─
SM 1550nm
SM 1310nm/OFF
─
AQ1205E
SM 1310nm
SM 1550nm/OFF
OFF
SM 1550nm
SM 1310nm/OFF
OFF
SM 1625nm
OFF
─
AQ1205F
SM 1310nm
SM 1550nm/OFF
OFF
SM 1550nm
SM 1310nm/OFF
OFF
SM 1650nm
OFF
─
* The AQ1200B and AQ1200C do not have Multi WL mode.