6-10
IM 701730-01E
Setting the Trigger Coupling
You can select from the following.
AC: Uses a signal that is obtained by removing the DC component from the trigger
source signal.
DC: Uses the trigger source signal as-is.
Setting the HF Rejection
Specify 15 kHz or 20 MHz if you wish to use a signal that is obtained by removing the
high frequency components (frequency components greater than 15 kHz or 20 MHz)
from the trigger source signal as the trigger source.
Setting the Hysteresis
Sets a width to the trigger level so that triggers are not activated by small changes in the
trigger signal.
:
Approximately 0.3 divisions* of hysteresis around the trigger level.
:
Approximately one division* of hysteresis around the trigger level.
*
The value above is an approximate value. It is not strictly warranted.
Setting the Hold Off
See section 6.4.
6.5 Setting the Edge Trigger (SIMPLE)