10-49
IM 701730-01E
Analyzing and Searching Waveforms
10
Measurement per Cycle and Statistical Processing within the Measurement Range
3.
Press the
Cycle Statistics
soft key.
4.
Perform the following setup procedures as necessary. For the operating procedure,
see section 10.6.
•
Selecting the measurement parameters (steps 4 to 8 on page 10-41).
•
Setting the measurement range (steps 20 and 21 on page 10-43).
•
Setting distal, mesial, and proximal values (steps 22 to 31 on page 10-43).
•
Setting high and low values (step 32 on page 10-44).
Selecting the Waveform for Determining the Cycle
5.
Press the
Cycle Trace
soft key. The Cycle Trace menu appears.
6.
Press the
Own
,
CH1
to
CH4(2)
,
Math1
, or
Math2
soft key to select the waveform
used to determine the cycle.
•
On the DL1735E/DL1740E/DL1740EL, you can select from Own, CH1 to CH4, Math1, and
Math2.
•
On the DL1720E, you can select Own, CH1, CH2, and Math1.
Executing/Aborting Measurement per Cycle and Statistical Processing within the
Measurement Range
7.
Press the
START/STOP
key to stop waveform acquisition.
8.
Press the
Measure Exec
soft key. Measurement per cycle and statistical
processing within the measurement range are performed on the displayed
waveform from the left end to the right end of the screen (from the oldest
waveform). The words Measure Exec change to Measure Abort.
To abort the measurement and statistical processing, press the
Measure Abort
soft key. The measurement and statistical processing are aborted, and the words
Measure Abort change to Measure Exec.
Statistical values display example
10.7 Statistical Processing of the Automated Measurement Values of Waveform Parameters