6-37
IM 701730-01E
Triggers
6
• For 480/60p: 8 to 2251
524 525
(1)
(2)
(3)
(4)
(5)
(6)
(7)
8
9
10
11
12
13
14
Line number
2207 2208
526
568
2209
2251
Line numbers inside the parentheses cannot be set.
43
30H
42H
6H
6H
Frame Skip
This function is used to skip frames such as when the color burst is inverted every frame.
You can select the number of frames at which this operation is repeated.
1: Activates a trigger at the specified field every time.
2: Skips 1 frames and activates a trigger at the specified field of the succeeding frame.
This operation is repeated every two frames.
4: Skips three frames and activates a trigger at the specified field of the succeeding
frame. This operation is repeated every four frames.
8: Skips seven frames and activates a trigger at the specified field of the succeeding
frame. This operation is repeated every eight frames.
Note
•
Video signals can only be input to CH1. All other channels do not support video signals.
•
Hold-off, trigger coupling, and HF rejection settings are ignored.
6.14 Setting the TV Trigger (ENHANCED)