10-67
IM 701730-01E
Analyzing and Searching Waveforms
10
9.
Use
jog shuttle & SELECT
to set the waveform acquisition count (ACQ Count
box).
10.
Use
jog shuttle & SELECT
to set the instrument action when the condition is met
(Action box).
Determination
parameter Param1
Determination
parameter Param2
Determination
parameter Param3
Determination
parameter Param4
Select whether to perform determination
using the corresponding determination
parameter (ON/OFF).
Select the determination target
waveform from CH1 to CH4 (2), Math1,
or Math2.
Select the measurement parameter.
Set the upper and lower limits used to
determine the true/false condition of the
measure values.
Select which condition when the
measured value exits the range defined
by the upper and lower limits (OUT) or
when the waveform enters the range (IN)
is considered true.
Set the determination logic to AND or
OR.
Set the determination operation to
Single or Continue.
Set the number of waveform
acquisitions.
Set the action taken by the instrument
when the condition is met.
11.
Press
ESC
. The Setup dialog box closes.
Setting the Determination Range
12.
Press the
T-Range1/T-Range2
soft key to set the jog shuttle control to T-Range1,
T-Range2, or both T-Range1 and T-Range 2.
•
If you select T-Range1, you can move T-Range1.
•
If you select T-Range2, you can move T-Range2.
•
If you select both T-Range1 and T-Range2, you can move T-Range1 and T-Range2
horizontally without changing the spacing between the two. The value of the digit being
specified by T-Range1 changes.
10.10 Performing GO/NO-GO Determination Using Automated Measurement of Waveform Parameters