4.5 Executing Auto Setup
Explanation
The auto setup function automatically sets the key settings such as V/div, T/div, and
trigger level that are appropriate for the input signal.
Center Position after Auto Setup
The center position after auto setup is 0 V.
Applicable Channels
Auto setup is performed on all channels.
Loaded Waveforms
After Auto Setup, loaded waveforms are unloaded. Press the Undo Auto Setup soft key
to recover unloaded waveforms.
Canceling Auto Setup
Pressing the Undo Auto Setup soft key sets the instrument back to the conditions that
existed immediately before auto setup. However, when you turn OFF the power switch,
the settings that existed immediately before auto setup are cleared. Therefore, the “Undo
Auto Setup” operation is not possible in this case.
Applicable Waveforms for Auto Setup
Frequency:
Approximately 50 Hz or higher
Absolute value of the input voltage:
Maximum value is greater than or equal to
approximately 20 mV (1:1)
Type:
Repetitive waveform (not complex)
Input coupling:
DC
Note
The auto setup function may not work properly if the waveform includes DC components or
high-frequency components.