6-29
IM 701730-01E
Triggers
6
Setting the Level, Hysteresis, Trigger Coupling, and HF Rejection
8.
Press the
Level/Coupling
soft key. The Level/Coupling dialog box opens.
The settings related to CH3 and CH4 are not displayed on the DL1720E.
9.
Like in the case of the A->B(N) trigger (see section 6.8), set the level, hysteresis,
trigger coupling, and HF rejection.
If Window is ON, set the window position and width, trigger coupling, and HF rejection.
The setting is the same as the window trigger. See section 6.13.
*1. Set the level of each channel.
*2. Select or for the hysteresis of each channel.
*3. Select DC or AC for the trigger coupling of each channel
*4. Select OFF, 200 MHz, or 15 kHz for the HR rejection of each channel.
*5. Set the position of the Window for each channel.
*6. Set the width of the Window for each channel.
*1
*2
*5
*6
*3
*4
When Window is
When Window is ON
10.
Press
ESC
. The Level/Coupling dialog box closes.
Note
The trigger level, hysteresis, trigger coupling, and HF rejection settings apply to both simple
and enhanced triggers.
Setting the Hold Off
11.
Set the hold off time according to the procedures given in section 6.4.
6.12 Setting the OR Trigger (ENHANCED)