5-13
IM 701730-01E
Vertical and Horizontal Axes
5
Explanation
The key settings of V/div, input coupling, trigger level, and other items are automatically
set to the optimum values (or arbitrary values) for the CMOS signal or ECL signal. You
can also automatically set them to the optimum values for the current probe 700937,
701930, 701931, or 701932 (sold separately).
You can set each channel separately or set all the channels to the same settings.
Setup after Executing Preset
Preset type
CMOS(5 V)
CMOS(3.3 V)
ECL
User
Input coupling
DC1 M
Ω
DC1 M
Ω
DC1 M
Ω
DC1 M
Ω
Trigger coupling
DC
DC
DC
DC
Probe
1:1, 10:1, 100:1, 1000:1, 10 A:1 V(0.1 V/A),100 A:1 V
Select from (0.01 V/A)
V/div
2 V/div
1 V/div
1
200 mV/div
1
Arbitrary
2
Offset voltage
0 V
0 V
–1.3 V
Arbitrary
2
Trigger level
2.5 V
1.65 V
–1.3 V
Arbitrary
2
1. 2 V/div (2 A/div) for 1000:1.
2. For the selectable ranges, see section 5.2, “Setting V/div,” 5.6, “Setting the Offset Voltage,” and
6.5, “Setting the Edge Trigger (SIMPLE).”