YL9130 Column Compartment
Chapter 1. Introduction
YL9130 column compartment is a thermostat to control column temperature of HPLC system. YL9130
column compartment uses RTD sensor to read temperature and PID control to control peltier device, it
provides wide temperature range and excellent temperature accuracy. The followings are standard
specification of YL9130 column compartment.
1-1. Specifications
1) Temperature range : 4
°
C (Cooling) - 90
°
C
2) Temperature stability :
±
0.05
°
C
3) Temperature accuracy :
±
0.5
°
C
4) Temperature precision :
±
0.1
°
C
5) Temperature programs : 40 Steps
6) Column capacity : Total 4 columns (max. 4ea of 300mm columns)
7) Column switching : One automatically 6-port valve ( optional )
8) Heat-up time : 16 minutes from 4
°
C to 90
°
C
9) Cool-down time : 13 minutes from 90
°
C to 4
°
C
10) Preheat : 200mm heat exchanger, OD 1/16 inch
Y
L9130
11) External input : Start, Ready
12) External output : Start, Ready Mark out
13) Communications : LAN
14) Safety & maintenance : Leak detection, Diagnostics, Error detection
15) Dimensions : 385 X 160 X 565mm (width X height X depth)
16) Line Voltage : 110 or 220 VAC,
±
10%
17) Line frequency : 50/60Hz,
±
5%
18) Power consumption : 150W
1
Summary of Contents for YL9100 HPLC
Page 1: ......
Page 2: ...YL9100 HPLC SYSTEM YL9101 VACUUM DEGASSER USER MANUAL...
Page 9: ...YL9100 HPLC SYSTEM YL9110 QUATERNARY SOLVENT DELIVERY PUMP USER MANUAL...
Page 44: ...YL9100 HPLC SYSTEM YL9111 BINARY PUMP USER MANUAL...
Page 79: ...YL9100 HPLC SYSTEM YL9112 ISOCRATIC PUMP USER MANUAL...
Page 114: ...YL9100 HPLC System YL9120 UV Vis Detector USER MANUAL...
Page 136: ...YL9100 HPLC SYSTEM YL9130 COLUMN COMPARTMENT USER MANUAL...
Page 150: ...YL9100 HPLC SYSTEM YL9160 PDA Detector USER MANUAL...
Page 169: ...YL9100 HPLC System YL9170 RI Detector USER MANUAL...
Page 188: ......