YL9112 Isocratic Pump
Chapter 5. Maintenance
In the event that problem occurs or it is necessary to change part due to wear of seal in using YL9112
isocratic pump, perform maintenance for instrument by referring to the following items.
5-1. Caution
In order to protect instrument, take care for the following items in using it.
1) After using solvent with sediment such as buffer solution, replace solvent with pure water at first
and then methanol or iso-propanol and make it flow for 30 minutes using each solvent at
1.0ml/min flow rate.
Y
L9112
2) Do not use solvent to corrode stainless steel material that is less than pH 2.3.
Material
Solvents to avoid
PEEK
Carbon Tetrachloride, Liquid Chlorine
Methylene Chloride, Tetrahydrofuran
Teflon(PTFE)
Dimethyl Formamide, Diethylamine
SS316 Phosphoric
Acid(Conc,
Rm
Temp)
3) Do not install instrument where corrosive gas is generated or where there is carpet on floor.
4) Do not change flow rate rapidly in order to prevent from wrong operation of instrument, damage
to column and damage to damper.
5) Do not operate instrument with excessive force.
5-2. Replacement of high pressure seal
If instrument is used for long time, high pressure seal is worn out to produce leakage of solution. In this
case, after replacing it with new seal, it is necessary to condition it in order to make seal be used for long
time at high pressure. Change and condition high pressure seal in the following method. In case
instrument is used for the first time after purchasing, it is desired to perform training and it is better to
23
Summary of Contents for YL9100 HPLC
Page 1: ......
Page 2: ...YL9100 HPLC SYSTEM YL9101 VACUUM DEGASSER USER MANUAL...
Page 9: ...YL9100 HPLC SYSTEM YL9110 QUATERNARY SOLVENT DELIVERY PUMP USER MANUAL...
Page 44: ...YL9100 HPLC SYSTEM YL9111 BINARY PUMP USER MANUAL...
Page 79: ...YL9100 HPLC SYSTEM YL9112 ISOCRATIC PUMP USER MANUAL...
Page 114: ...YL9100 HPLC System YL9120 UV Vis Detector USER MANUAL...
Page 136: ...YL9100 HPLC SYSTEM YL9130 COLUMN COMPARTMENT USER MANUAL...
Page 150: ...YL9100 HPLC SYSTEM YL9160 PDA Detector USER MANUAL...
Page 169: ...YL9100 HPLC System YL9170 RI Detector USER MANUAL...
Page 188: ......