YL9120 UV/Vis Detector
In the table below, edit programming of time table, and setup the detector status during idle state
Time Constant : To input the value of digital filter.
If value is larger, the baseline is smoother while peak width is wider.
Auto Zero : To change the absorbance to zero
Auto Offset : To prolong the baseline by adjustment automatically, If either wavelength or polarity is
changed.
Data Filter : To select which filter you use.
If the signal is opposite when you use with another device, change it on the External In/Out.
If you use an A/D converter for data acquisition, setup the each value of source, type, conversion factor
and offset.
12