YL9111 Binary Pump
•
Pump Head Assembly
Pump head is real working part to deliver solvent by piston movement and check valve. It consists of
plunger, check valve, high pressure seal, low pressure seal and rinsing port. YL9111 Binary pump built
in automatic rinsing port to clean the head assembly efficiently if uses buffer solution.
•
Auto Rinsing Pump
Auto rinsing pump delivers cleaning solvent from the rinsing solvent bottle to the inside of pump head.
YL9111 Binary pump rinse the system every 3 minutes automatically.
•
Prime/purge valve
This valve is used for priming the pump. Fill the solvent inside of tubing from the solvent bottle if you use
the pump for the first time or the tubing lines are empty. Remove a plug on the prime/purge valve and
suck the solvent using syringe, or click “Prime Start” button on the software to operate the micro pump.
•
Prime micro pump
Using this pump, the pump circulates rinse solvent into the pump head.
•
Mixer
It is static mixer to improve mixing efficiency of solvent.
•
Pulse damper
YL9111 Binary pump reduces pulse from the cam operation by the diaphragm damper. YL9111 Binary
pump provides constant and pulseless flow using compressibility compensation and pulse damper, so
the detector that affected by flow stability can be used with YL9111 pump.
•
Pressure transducer
It checks real time system pressure. The pump uses this pressure to protect system and to operate
compressibility compensation and even compensation. YL9111 Binary pump uses continuous flow path
type pressure sensor.
•
In-line filter
It removes fine particles that are not filtered by solvent filter or made by worn of high pressure seal.
4
Summary of Contents for YL9100 HPLC
Page 1: ......
Page 2: ...YL9100 HPLC SYSTEM YL9101 VACUUM DEGASSER USER MANUAL...
Page 9: ...YL9100 HPLC SYSTEM YL9110 QUATERNARY SOLVENT DELIVERY PUMP USER MANUAL...
Page 44: ...YL9100 HPLC SYSTEM YL9111 BINARY PUMP USER MANUAL...
Page 79: ...YL9100 HPLC SYSTEM YL9112 ISOCRATIC PUMP USER MANUAL...
Page 114: ...YL9100 HPLC System YL9120 UV Vis Detector USER MANUAL...
Page 136: ...YL9100 HPLC SYSTEM YL9130 COLUMN COMPARTMENT USER MANUAL...
Page 150: ...YL9100 HPLC SYSTEM YL9160 PDA Detector USER MANUAL...
Page 169: ...YL9100 HPLC System YL9170 RI Detector USER MANUAL...
Page 188: ......