YL9160 PDA Detector
Chapter 2. Installation
2-1. Inspection and site preparation
Carefully unpack the detector from the shipping box and inspect both the unit and packing for any
signs of damage. If any damage is noted, contact the shipping company immediately. In addition to this
manual, the shipping box contains a power cord, and any options which you ordered. Carefully check the
packing list against the contents of the container. If anything is missing, check the packing materials
carefully for the overlooked items. If items are missing, contact us or your supplier. Place the detector on
the bench where it will be used and familiarize yourself with the location and function of the controls and
connections.
Site requirement of YL9160 PDA Detector
1) Room with 20
℃
temperature with variation ± 5
℃
with and 60% humidity
2) Where no direct and straight sunshine
3) Plain floor without carpet
4) Where having spare space of 20cm or more
5) Where there are 10% or less voltage change, no frequency change and 100
Ω
or less grounding
point
6) Where there is no generation of corrosive gas and ventilation is well done
7) Where stable power of 110 or 220V is supplied,
±
10%
8) Where not receiving electromagnetic induction from large transformer, high frequency heater,
UPS, etc.
Place the detector on a laboratory benchtop in close proximity to the HPLC column outlet. Allow at least
5 inches of clear space between the rear panel of the unit and any wall or obstruction. This provides both
access to the rear panel connections and a free flow of air
In addition to the detector itself, you will need the following items for setup and initial operation :
1) YL-Clarity software or Chromatograph Data System.
2) Pump
3) Column
4) Standard test mixture
4
Summary of Contents for YL9100 HPLC
Page 1: ......
Page 2: ...YL9100 HPLC SYSTEM YL9101 VACUUM DEGASSER USER MANUAL...
Page 9: ...YL9100 HPLC SYSTEM YL9110 QUATERNARY SOLVENT DELIVERY PUMP USER MANUAL...
Page 44: ...YL9100 HPLC SYSTEM YL9111 BINARY PUMP USER MANUAL...
Page 79: ...YL9100 HPLC SYSTEM YL9112 ISOCRATIC PUMP USER MANUAL...
Page 114: ...YL9100 HPLC System YL9120 UV Vis Detector USER MANUAL...
Page 136: ...YL9100 HPLC SYSTEM YL9130 COLUMN COMPARTMENT USER MANUAL...
Page 150: ...YL9100 HPLC SYSTEM YL9160 PDA Detector USER MANUAL...
Page 169: ...YL9100 HPLC System YL9170 RI Detector USER MANUAL...
Page 188: ......